Abstract
Rehabilitation after laryngectomy includes more and more the use of airflow-regulating devices such as shunt valves (SVs), tracheostoma valves (TSVs), and heat and moisture exchange (HME) filters. In determining the quality of those devices, airflow resistance is a very important factor. It is currently defined as pressure drop divided by airflow. However, for most applications, this definition does not result in a pressure- and airflow-independent parameter. Therefore, a new set of parameters is defined and applied to pressure-airflow curves of airflow-regulating devices. Pressure drop over TSVs and HME filters appears to have a squared relationship with flow. In SVs, it has a linear relationship. The new set of parameters describes the pressure-airflow relationship properly for all considered devices. In conclusion, theoretical predictions of flow mechanics appear to be valid for SVs, TSVs, and HME filters. Only 2 coefficients are necessary to describe the pressure-flow characteristics of these airflow-regulating devices, independent of pressure drop over and flow through the device.
Original language | English |
---|---|
Pages (from-to) | 639-645 |
Number of pages | 7 |
Journal | Annals of otology, rhinology and laryngology |
Volume | 110 |
Issue number | 7 I |
DOIs | |
Publication status | Published - 1 Jan 2001 |
Externally published | Yes |
Keywords
- Airflow resistance
- Heat and moisture exchange filter
- Laryngectomy
- Pressure drop
- Shunt valve
- Tracheostoma valve