Bioresorbable Polymer-Coated Orsiro Versus Durable Polymer-Coated Resolute Onyx Stents (BIONYX): Rationale and design of the randomized TWENTE IV multicenter trial

Liefke C. van der Heijden, Marlies M. Kok, Gillian A.J. Jessurun, Carl E. Schotborgh, Ariel Roguin, Edouard Benit, Adel Aminian, Peter W. Danse, Marije M. Löwik, Gerard C.M. Linssen, Job van der Palen, Carine J.M. Doggen, Clemens von Birgelen* (Corresponding Author), Paolo Zocca

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

10 Citations (Scopus)
229 Downloads (Pure)

Abstract

Aim: The aim was to compare in a noninferiority trial the efficacy and safety of 2 contemporary drug-eluting stents (DESs): a novel, durable polymer-coated stent versus an established bioabsorbable polymer-coated stent.

Methods and results: The BIONYX trial (ClinicalTrials.gov-no.NCT02508714) is an investigator-initiated, prospective, randomized, patient- and assessor-blinded, international, multicenter study in all-comer patients with all types of clinical syndromes and lesions who require percutaneous coronary interventions with DES. Patients at 7 study sites in the Netherlands, Belgium, and Israel were randomly assigned (1:1, stratified for gender and diabetes mellitus) to treatment with the novel, zotarolimus-eluting, durable polymer-coated Resolute Onyx stent that has a radiopaque, thin-strut, CoreWire stent platform versus the sirolimus-eluting, bioresorbable polymer-coated Orsiro stent (reference device) that has a very thin-strut, cobalt-chromium stent backbone. The primary end point is the 1-year incidence of the composite clinical end point target vessel failure consisting of cardiac death, target vessel–related myocardial infarction, or clinically indicated target vessel revascularization. A power calculation, assuming a target vessel failure rate of 6.0% (noninferiority margin 2.5%), revealed that 2,470 study patients would give the study 80% power (α level 5%), allowing for up to 3% loss to follow-up. The first patient was enrolled on October 7, 2015; on December 23, 2016, the last patient entered the study.

Conclusions: BIONYX is a large-scale, prospective, randomized, international, multicenter trial comparing a novel DES with durable coating versus a reference DES with biodegradable coating in all-comers. The study is the first randomized assessment of the Resolute Onyx stent, which is an often-used DES outside the United States.

Original languageEnglish
Pages (from-to)25-32
Number of pages8
JournalAmerican heart journal
Volume198
DOIs
Publication statusPublished - 1 Apr 2018