Abstract
Purpose: Worldwide, 600,000 infants are born extremely premature (EP) each year. Supporting the immature lungs of these smallest patients today using mechanical ventilation can cause severe lung damage. Extracorporeal gas exchange (ECMO) used as an “artificial placenta” is expected to provide an alternative treatment to allow full maturation of the lungs. The long-term application for patients doubling their weight every six weeks, requires a new type of “growing” oxygenator.
Materials & Methods: We developed a gas-exchange-area-adjustable oxygenator that allows to operate at two different size stages without changing the device nor increasing the flow resistance. Our prototype houses two concentric oxygenator-chambers of equal gas exchange area. When the outer chamber is in operation, the oxygenator provides the required gas exchange area for a 24-week infant. Single-chamber operation allows adaequate gas exchange of volume flows from 50 - 125 ml/min while double-chamber operation then covers flow-demands from 125 - 200 ml/min. Both operational modes in combination can cover the needs of infants from 24 to 28 weeks postmenstrual age. We performed gas-transfer measurements of the prototype in accordance with ISO 7199 and FDA guideline.
Results: Gas transport performance for the outer cylinder operating at Vblood = 50 mL/min (24 week premature infant) is VO2 = 1.8 mL/min (= 36 mLO2/Lblood flow). At a flow of Vblood = 125 mL/min the second chamber was switched on and the gas transfer performance increased by 50.1 %. Both chambers have a gas transport performance of VO2 = 4.19 ml/min at Vblood = 150 ml/min (= 28 mLO2/Lblood flow)
Conclusion: It could be demonstrated that the gas-transfer of the newly proposed volume-adjustable oxygenator design is sufficient to increase the gas transfer performance while increasing the blood volume flow. We consider this a first milestone for oxygenation of growing EP infants without the need to replace the extracorporeal circuit, avoiding the associated risks.
Materials & Methods: We developed a gas-exchange-area-adjustable oxygenator that allows to operate at two different size stages without changing the device nor increasing the flow resistance. Our prototype houses two concentric oxygenator-chambers of equal gas exchange area. When the outer chamber is in operation, the oxygenator provides the required gas exchange area for a 24-week infant. Single-chamber operation allows adaequate gas exchange of volume flows from 50 - 125 ml/min while double-chamber operation then covers flow-demands from 125 - 200 ml/min. Both operational modes in combination can cover the needs of infants from 24 to 28 weeks postmenstrual age. We performed gas-transfer measurements of the prototype in accordance with ISO 7199 and FDA guideline.
Results: Gas transport performance for the outer cylinder operating at Vblood = 50 mL/min (24 week premature infant) is VO2 = 1.8 mL/min (= 36 mLO2/Lblood flow). At a flow of Vblood = 125 mL/min the second chamber was switched on and the gas transfer performance increased by 50.1 %. Both chambers have a gas transport performance of VO2 = 4.19 ml/min at Vblood = 150 ml/min (= 28 mLO2/Lblood flow)
Conclusion: It could be demonstrated that the gas-transfer of the newly proposed volume-adjustable oxygenator design is sufficient to increase the gas transfer performance while increasing the blood volume flow. We consider this a first milestone for oxygenation of growing EP infants without the need to replace the extracorporeal circuit, avoiding the associated risks.
Original language | English |
---|---|
Pages (from-to) | 59-59 |
Journal | ASAIO Journal |
Volume | 69 |
Issue number | Suppl. 2 |
DOIs | |
Publication status | Published - 1 Jun 2023 |
Event | ASAIO 68th Annual Conference 2023 - San Francisco Marriott Marquis, San Francisco, United States Duration: 14 Jun 2023 → 17 Jun 2023 Conference number: 68 |