TY - JOUR
T1 - Cardiorespiratory monitoring with a wireless and nonadhesive belt measuring diaphragm activity in preterm and term infants
T2 - A multicenter non-inferiority study
AU - Scholten, Anouk W.J.
AU - Zhan, Zhuozhao
AU - Niemarkt, Hendrik J.
AU - Vervoorn, Marieke
AU - van Leuteren, Ruud W.
AU - de Jongh, Frans H.
AU - van Kaam, Anton H.
AU - Heuvel, Edwin R.van den
AU - Hutten, Gerard J.
N1 - Publisher Copyright:
© 2023 The Authors. Pediatric Pulmonology published by Wiley Periodicals LLC.
PY - 2023/12
Y1 - 2023/12
N2 - Introduction: We determined if the heart rate (HR) monitoring performance of a wireless and nonadhesive belt is non-inferior compared to standard electrocardiography (ECG). Secondary objective was to explore the belt's respiratory rate (RR) monitoring performance compared to chest impedance (CI). Method: In this multicenter non-inferiority trial, preterm and term infants were simultaneously monitored with the belt and conventional ECG/CI for 24 h. HR monitoring performance was estimated with the HR difference and ability to detect cardiac events compared to the ECG, and the incidence of HR-data loss per second. These estimations were statistically compared to prespecified margins to confirm equivalence/non-inferiority. Exploratory RR analyses estimated the RR trend difference and ability to detect apnea/tachypnea compared to CI, and the incidence of RR-data loss per second. Results: Thirty-nine infants were included. HR monitoring with the belt was non-inferior to the ECG with a mean HR difference of 0.03 beats per minute (bpm) (standard error [SE] = 0.02) (95% limits of agreement [LoA]: [−5 to 5] bpm) (p < 0.001). Second, sensitivity and positive predictive value (PPV) for cardiac event detection were 94.0% (SE = 0.5%) and 92.6% (SE = 0.6%), respectively (p ≤ 0.001). Third, the incidence of HR-data loss was 2.1% (SE = 0.4%) per second (p < 0.05). The exploratory analyses of RR showed moderate trend agreement with a mean RR-difference of 3.7 breaths/min (SE = 0.8) (LoA: [−12 to 19] breaths/min), but low sensitivities and PPV's for apnea/tachypnea detection. The incidence of RR-data loss was 2.2% (SE = 0.4%) per second. Conclusion: The nonadhesive, wireless belt showed non-inferior HR monitoring and a moderate agreement in RR trend compared to ECG/CI. Future research on apnea/tachypnea detection is required.
AB - Introduction: We determined if the heart rate (HR) monitoring performance of a wireless and nonadhesive belt is non-inferior compared to standard electrocardiography (ECG). Secondary objective was to explore the belt's respiratory rate (RR) monitoring performance compared to chest impedance (CI). Method: In this multicenter non-inferiority trial, preterm and term infants were simultaneously monitored with the belt and conventional ECG/CI for 24 h. HR monitoring performance was estimated with the HR difference and ability to detect cardiac events compared to the ECG, and the incidence of HR-data loss per second. These estimations were statistically compared to prespecified margins to confirm equivalence/non-inferiority. Exploratory RR analyses estimated the RR trend difference and ability to detect apnea/tachypnea compared to CI, and the incidence of RR-data loss per second. Results: Thirty-nine infants were included. HR monitoring with the belt was non-inferior to the ECG with a mean HR difference of 0.03 beats per minute (bpm) (standard error [SE] = 0.02) (95% limits of agreement [LoA]: [−5 to 5] bpm) (p < 0.001). Second, sensitivity and positive predictive value (PPV) for cardiac event detection were 94.0% (SE = 0.5%) and 92.6% (SE = 0.6%), respectively (p ≤ 0.001). Third, the incidence of HR-data loss was 2.1% (SE = 0.4%) per second (p < 0.05). The exploratory analyses of RR showed moderate trend agreement with a mean RR-difference of 3.7 breaths/min (SE = 0.8) (LoA: [−12 to 19] breaths/min), but low sensitivities and PPV's for apnea/tachypnea detection. The incidence of RR-data loss was 2.2% (SE = 0.4%) per second. Conclusion: The nonadhesive, wireless belt showed non-inferior HR monitoring and a moderate agreement in RR trend compared to ECG/CI. Future research on apnea/tachypnea detection is required.
KW - newborn infants
KW - preterm infants
KW - transcutaneous electromyography
UR - https://www.scopus.com/pages/publications/85173517544
U2 - 10.1002/ppul.26695
DO - 10.1002/ppul.26695
M3 - Article
AN - SCOPUS:85173517544
SN - 8755-6863
VL - 58
SP - 3574
EP - 3581
JO - Pediatric pulmonology
JF - Pediatric pulmonology
IS - 12
ER -