Abstract
To determine the rheobase and the chronaxie of excitable cells from strength–duration curves both constant-current pulses and constant-voltage pulses are applied. Since the complex impedance of the electrode-tissue interface varies with both the pulsewidth and the stimulation voltage, chronaxie values estimated from voltage–duration measurements will differ from the proper values as determined from current–duration measurements. To allow a comparison of chronaxie values obtained by the two stimulation methods, voltage–duration curves were measured in human subjects with a deep brain stimulation electrode implanted, while the current and the load impedance of the stimulation circuit were determined in vitro as a function of both stimulation voltage and pulsewidth. Chronaxie values calculated from voltage–duration data were shown to be 30–40% below those estimated from current–duration data. It was also shown that in the normal range of stimulation amplitudes (up to 7 V) the load impedance increases almost linearly with the pulsewidth. This result led us to present a simple method to convert voltage–duration data into current–duration data, thereby reducing the error in the calculated chronaxie values to ≈6%. For this purpose voltage–duration data have to be measured for pulses up to 10–20 times the expected chronaxie.
Original language | Undefined |
---|---|
Pages (from-to) | 45-50 |
Number of pages | 6 |
Journal | Journal of neuroscience methods |
Volume | 2000 |
Issue number | 97 |
DOIs | |
Publication status | Published - 2000 |
Keywords
- METIS-111770
- Rheobase
- Deep Brain Stimulation
- Current–duration measurements
- IR-74311
- Chronaxie
- Voltage–duration measurements