Comparison of capillary zone electrophoresis performance of powder-blasted and hydrogen fluoride-etched microchannels in glass

Research output: Contribution to journalArticleAcademicpeer-review

37 Citations (Scopus)
5 Downloads (Pure)

Abstract

The applicability of glass chips with powder-blasted microchannels for electrophoretic separations was examined, and the performance was compared to microchannels etched with hydrogen fluoride (HF), using bicarbonate buffer and rhodamine B and fluorescein as model compounds. The measured electroosmotic mobilities in all chips were comparable, with values of ca. 7×10-4 cm2 V-1s-1. The effect of electrical field strength and detection length on the separation efficiency was monitored. It was found that the main source of dispersion is of the Taylor-Aris type, which was discussed in relation to channel roughness differences. Although in powder-blasted channels with a separation length of 8.20 cm, 7-9 times lower plate numbers were obtained than in a HF-etched channel with similar dimensions, successful separation of five fluorescein isothiocyanate (FITC)-labeled amino acids was obtained on a powder-blasted chip within 80 s. Efficiencies of up to 360 000 plates/m were demonstrated on this chip, when a higher buffer concentration was used at a field strength of 664 V/cm. It can be concluded that powder-blasted microchannel chips, although they have a lower separation efficiency compared to HF-etched chips, perform well enough for many applications. Powder blasting can therefore be considered a low-cost and efficient alternative to HF etching, in particular because of the possibility to fabricate access holes through the glass with the same process.
Original languageUndefined
Pages (from-to)162-171
Number of pages10
JournalElectrophoresis
Volume24
Issue number1-2
DOIs
Publication statusPublished - 2003

Keywords

  • METIS-214212
  • IR-46181
  • Glass chips
  • Capillary zone electrophoresis
  • Lab on a chip

Cite this