TY - JOUR
T1 - Construct Validity and Reproducibility of Handheld Ultrasound Devices in Carotid Artery Diameter Measurement
AU - Willems, Loes
AU - Vermeulen, Jenske
AU - Wiegerinck, Anouk
AU - Fekkes, Stein
AU - Reijnen, Michel
AU - Warlé, Michiel
AU - De Korte, Chris
AU - Thijssen, Dick
N1 - Funding Information:
The data generated during the present study are available from the corresponding author on reasonable request.
Publisher Copyright:
© 2022 The Authors
PY - 2023/3
Y1 - 2023/3
N2 - The construct validity and reproducibility of three commonly used handheld ultrasound (US) devices in measuring carotid arterial diameter was evaluated: Telemed MicrUs EXT-1H (Telemed, Vilnius, Lithuania), Butterfly iQ (Butterfly Network, Inc., Guilford, CT, USA) and Philips Lumify (Philips Healthcare, Best, The Netherlands). An in vitro setup was built to evaluate construct validity, compared with high-end US, and intra-observer variability of handheld US devices. Handheld devices showed a mean difference of 0.023 ± 0.030 cm, 0.012 ± 0.037 cm and 0.009 ± 0.046 cm for, respectively, Telemed, Butterfly and Lumify in comparison with high-end US devices. Intraclass agreement with the high-end system as well as intra-observer variability for handheld US devices was classified as excellent, with all values greater than 0.95. Subsequently, inter-observer variability of handheld US devices was investigated in an in vivo setup with 20 healthy volunteers. Inter-observer variability was classified as excellent for Telemed (0.901), good for Lumify (0.827) and moderate for Butterfly (0.684) with a difference of, respectively, 0.005 ± 0.031 cm, 0.020 ± 0.050 cm and –0.003 ± 0.033 cm. In conclusion, handheld US devices demonstrated an excellent construct validity and intra-observer variability. Additionally, excellent-to-good inter-observer variability for Telemed and Lumify was observed, and Butterfly demonstrated a moderate inter-observer agreement. These results indicate that handheld US devices are effective for measuring carotid arterial diameter.
AB - The construct validity and reproducibility of three commonly used handheld ultrasound (US) devices in measuring carotid arterial diameter was evaluated: Telemed MicrUs EXT-1H (Telemed, Vilnius, Lithuania), Butterfly iQ (Butterfly Network, Inc., Guilford, CT, USA) and Philips Lumify (Philips Healthcare, Best, The Netherlands). An in vitro setup was built to evaluate construct validity, compared with high-end US, and intra-observer variability of handheld US devices. Handheld devices showed a mean difference of 0.023 ± 0.030 cm, 0.012 ± 0.037 cm and 0.009 ± 0.046 cm for, respectively, Telemed, Butterfly and Lumify in comparison with high-end US devices. Intraclass agreement with the high-end system as well as intra-observer variability for handheld US devices was classified as excellent, with all values greater than 0.95. Subsequently, inter-observer variability of handheld US devices was investigated in an in vivo setup with 20 healthy volunteers. Inter-observer variability was classified as excellent for Telemed (0.901), good for Lumify (0.827) and moderate for Butterfly (0.684) with a difference of, respectively, 0.005 ± 0.031 cm, 0.020 ± 0.050 cm and –0.003 ± 0.033 cm. In conclusion, handheld US devices demonstrated an excellent construct validity and intra-observer variability. Additionally, excellent-to-good inter-observer variability for Telemed and Lumify was observed, and Butterfly demonstrated a moderate inter-observer agreement. These results indicate that handheld US devices are effective for measuring carotid arterial diameter.
KW - Carotid arteries
KW - Diameter
KW - Handheld ultrasound
KW - Ultrasonography
UR - http://www.scopus.com/inward/record.url?scp=85147113379&partnerID=8YFLogxK
U2 - 10.1016/j.ultrasmedbio.2022.11.013
DO - 10.1016/j.ultrasmedbio.2022.11.013
M3 - Article
C2 - 36567191
AN - SCOPUS:85147113379
SN - 0301-5629
VL - 49
SP - 866
EP - 874
JO - Ultrasound in medicine and biology
JF - Ultrasound in medicine and biology
IS - 3
ER -