N2 - By analysis of some important process variables, criteria for successful separation by continuous electrophoresis and electrochromatography in packed beds are derived. A general theory correlating power input, residence time and temperature rise in cylindrical and rectangular geometries is presented. The limitation of the separating capacity by transverse diffusion effects is shown to be predictable in terms of other operational conditions. These separation criteria appear to be in agreement with experimental evidence, and may find analytical as well as preparative application.
AB - By analysis of some important process variables, criteria for successful separation by continuous electrophoresis and electrochromatography in packed beds are derived. A general theory correlating power input, residence time and temperature rise in cylindrical and rectangular geometries is presented. The limitation of the separating capacity by transverse diffusion effects is shown to be predictable in terms of other operational conditions. These separation criteria appear to be in agreement with experimental evidence, and may find analytical as well as preparative application.