Cross-sectional changes in lung volume measured by electrical impedance tomography are representative for the whole lung in ventilated preterm infants

Pauline S. van der Burg, Martijn Miedema, Franciscus H.C. de Jongh, Inez Frerichs, Anton H. van Kaam

Research output: Contribution to journalArticleAcademicpeer-review

41 Citations (Scopus)
14 Downloads (Pure)

Abstract

Objective: Electrical impedance tomography measures lung volume in a cross-sectional slice of the lung. Whether these cross-sectional volume changes are representative of the whole lung has only been investigated in adults, showing conflicting results. This study aimed to compare cross-sectional and whole lung volume changes using electrical impedance tomography and respiratory inductive plethysmography. Design: A prospective, single-center, observational, nonrandomized study. Setting: The study was conducted in a neonatal ICU in the Netherlands. Patients: High-frequency ventilated preterm infants with respiratory distress syndrome. Interventions: Cross-sectional and whole lung volume changes were continuously and simultaneously measured by, respectively, electrical impedance tomography and respiratory inductive plethysmography during a stepwise recruitment procedure. End-expiratory lung volume changes were assessed by mapping the inflation and deflation limbs using both the pressure/impedance and pressure/inductance pairs and characterized by calculating the inflection points. In addition, oscillatory tidal volume changes were assessed at each pressure step. Measurements and Main Results: Twenty-three infants were included in the study. Of these, eight infants had to be excluded because the quality of the registration was insufficient for analysis (two electrical impedance tomography and six respiratory inductive plethysmography). In the remaining 15 infants (gestational age 28.0 ± 2.6 wk; birth weight 1,027 ± 514 g), end-expiratory lung volume changes measured by electrical impedance tomography were significantly correlated to respiratory inductive plethysmography measurements in 12 patients (mean r = 0.93 ± 0.05). This was also true for the upper inflection point on the inflation (r = 0.91, p < 0.01) and deflation limb (r = 0.83, p < 0.01). In 13 patients, impedance and inductance data also correlated significantly on oscillatory tidal volume/pressure relationships (mean r = 0.81 ± 0.18). Conclusions: This study shows that cross-sectional lung volume changes measured by electrical impedance tomography are representative for the whole lung and that this concept also applies to newborn infants.
Original languageUndefined
Pages (from-to)1524-1530
JournalCritical care medicine
Volume42
Issue number6
DOIs
Publication statusPublished - 2014

Keywords

  • METIS-316161
  • IR-99693

Cite this