Electrocardiogram-gated intravascular ultrasound image acquisition after coronary stent deployment facilitates on-line three-dimensional reconstruction and automated lumen quantification

Clemens von Birgelen, Gary S. Mintz, Antonino Nicosia, David P. Foley, Wim J. van der Giessen, Nico Bruining, Sergei G. Airiian, Jos R.T.C. Roelandt, Pim J. de Feyter, Patrick W. Serruys* (Corresponding Author)

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

82 Citations (Scopus)

Abstract

Objectives. This study evaluates the feasibility, reliability and reproducibility of electrocardiogram (ECG)-gated intravascular ultrasound (IVUS) image acquisition during automated transducer withdrawal and automated three-dimensional (3D) boundary detection for assessing on-line the result of coronary stenting. Background. Systolic-diastolic image artifacts frequently limit the clinical applicability of such automated analysis systems. Methods. In 30 patients, after successful angiography-guided implantation of 34 stents in 30 target lesions, we carried out IVUS examinations on-line with the use of ECG-gated automated 3D analyses and conventional manual analyses of two- dimensional images from continuous pullbacks. These on-line measurements were compared with off-line 3D reanalyses. The adequacy of stent deployment was determined by using ultrasound criteria for stent apposition, symmetry and expansion. Results. Gated image acquisition was successfully performed in all patients to allow on-line 3D analysis within 8.7 ± 0.6 min (mean ± SD). Measurements by on-line and off-line 3D analyses correlated closely (r ≤ 0.95), and the minimal stent lumen differed only minimally (8.6 ± 2.8 mm2 vs. 8.5 ± 2.8 mm2, p = NS). The conventional analysis significantly overestimated the minimal stent lumen (9.0 ± 2.7 mm2, p < 0.005) in comparison with results of both 3D analyses. Fourteen stents (41%) failed to meet the criteria by both 3D analyses, all of these not reaching optimal expansion, but only 7 (21%) were detected by conventional analysis (p < 0.02). Intraobserver and interobserver comparison of stent lumen measurements by the automated approach revealed minimal differences (0.0 ± 0.2 mm2 and 0.0 ± 0.3 mm2) and excellent correlations (r = 0.99 and 0.98, respectively). Conclusion. ECG-gated image acquisition after coronary stent deployment is feasible, permits on-line automated 3D reconstruction and analysis and provides reliable and reproducible measurements; these factors facilitate detection of the minimal lumen site.

Original languageEnglish
Pages (from-to)436-443
Number of pages8
JournalJournal of the American College of Cardiology
Volume30
Issue number2
DOIs
Publication statusPublished - 1 Aug 1997
Externally publishedYes

Fingerprint

Dive into the research topics of 'Electrocardiogram-gated intravascular ultrasound image acquisition after coronary stent deployment facilitates on-line three-dimensional reconstruction and automated lumen quantification'. Together they form a unique fingerprint.

Cite this