Electrode contact configuration and energy consumption in spinal cord stimulation

Cecilia C. de Vos*, Marjolein P. Hilgerink, Hendrik P.J. Buschman, Jan Holsheimer

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

11 Citations (Scopus)
16 Downloads (Pure)

Abstract

Objective: To test the hypothesis that in spinal cord stimulation, an increase in the number of cathodes increases the energy per pulse, contrary to an increase in the number of anodes, which decreases energy consumption per pulse.

Methods: Patients with an Itrel III (7425; Medtronic, Inc., Minneapolis, MN) implantable pulse generator and a Pisces-Quad (3487A; Medtronic, Inc.) implantable quadripolar lead were selected for this study. A set of 7 standard contact configurations was used for each patient. Resistor network models mimicking these configurations were constructed. The University of Twente’s Spinal Cord Stimulation software was used to simulate the effect of these contact configurations on large spinal nerve fibers. To allow a comparison of the measured and modeled energy per pulse, all values were normalized.

Results: Both the empirical and the modeling results showed an increase in energy consumption with an increasing number of cathodes. Although the patient data with 1 and 2 cathodes did not differ significantly, energy consumption was significantly higher when 3 cathodes were used instead of 1 or 2 cathodes. The average energy consumption was significantly higher when bipolar stimulation was used instead of monopolar cathodal stimulation. An increasing number of anodes caused a decrease in energy consumption.

Conclusion: When the paresthesia area can be covered with several configurations, it will be beneficial for the patient to program a configuration with 1 cathode and either no or multiple anodes.
Original languageEnglish
Pages (from-to)ons210-ons217
Number of pages8
JournalNeurosurgery
Volume65
Issue number6
DOIs
Publication statusPublished - 1 Dec 2009

Keywords

  • BSS-Electrical Neurostimulation

Fingerprint

Dive into the research topics of 'Electrode contact configuration and energy consumption in spinal cord stimulation'. Together they form a unique fingerprint.
  • Contact configuration and energy consumption in spinal cord stimulation

    de Vos, C. C., Hilgerink, M. P., Buschman, H. P. J. & Holsheimer, J., 2009, 4th European Conference of the International Federation for Medical and Biological Engineering: ECIFMBE 2008 23–27 November 2008 Antwerp, Belgium. Vander Sloten, J., Verdonck, P., Nyssen, M. & Haueisen, J. (eds.). Berlin, Heidelberg: Springer, p. 2503-2506 (IFMBE proceedings; vol. 22).

    Research output: Chapter in Book/Report/Conference proceedingChapterAcademic

    1 Downloads (Pure)

Cite this