Evaluation of flexible endoscope steering using haptic guidance

Rob Reilink, Stefano Stramigioli, Astrid M.L. Kappers, Sarthak Misra

    Research output: Contribution to journalArticleAcademicpeer-review

    26 Citations (Scopus)
    14 Downloads (Pure)

    Abstract

    Background - Steering the tip of a flexible endoscope relies on the physician’s dexterity and experience. For complex flexible endoscopes, conventional controls may be inadequate. Methods - A steering method based on a multi-degree-of-freedom haptic device is presented. Haptic cues are generated based on the endoscopic images. The method is compared against steering using the same haptic device without haptic cues, and against conventional steering. Human-subject studies were conducted in which 12 students and 6 expert gastroenterologists participated. Results - Experts are significantly faster when using the conventional method compared with using the haptic device, either with or without haptic cues. However, it is expected that the performance of the subjects with the haptic device will increase with experience. Conclusions - Using a haptic device may be a viable alternative to the conventional method for the control of complex flexible endoscopes. The results suggest that the use of haptic cues may reduce the patient discomfort
    Original languageUndefined
    Pages (from-to)178-186
    Number of pages9
    JournalInternational journal of medical robotics and computer assisted surgery
    Volume7
    Issue number2
    DOIs
    Publication statusPublished - 2011

    Keywords

    • Medical robotics
    • endoscope steering
    • haptic guidance
    • Image-guided surgery
    • METIS-278709
    • IR-76581
    • Colonoscopy
    • EWI-19947