Abstract
Objectives: To assess the feasibility of adding a tablet device inside the scanner room to assist needle-guide alignment during magnetic resonance (MR)-guided transrectal prostate biopsy. Methods: Twenty patients with one cancer-suspicious region (CSR) with PI-RADS score ≥ 4 on diagnostic multiparametric MRI were prospectively enrolled. Two orthogonal scan planes of an MR fluoroscopy sequence (~3 images/s) were aligned to the CSR and needle-guide pivoting point. Targeting was achieved by manipulating the needle-guide under MR fluoroscopy feedback on the in-room tablet device. Technical feasibility and targeting success were assessed. Complications and biopsy procedure times were also recorded. Results: Needle-guide alignment with the in-room tablet device was technically successful in all patients and allowed sampling after a single alignment step in 19/20 (95%) CSRs (median size 14 mm, range: 4-45). Biopsy cores contained cancer in 18/20 patients. There were no per-procedural or post-biopsy complications. Using the tablet device, the mean time to first biopsy was 5.8 ± 1.0 min and the mean total procedure time was 23.7 ± 4.1 min. Conclusions: Use of an in-room tablet device to assist needle-guide alignment was feasible and safe during MR-guided transrectal prostate biopsy. Initial experience indicates potential for procedure time reduction. Key Points: • Performing MR-guided prostate biopsy using an in-room tablet device is feasible. • CSRs could be sampled after a single alignment step in 19/20 patients. • The mean procedure time for biopsy with the tablet device was 23.7 min.
Original language | English |
---|---|
Pages (from-to) | 4824-4831 |
Number of pages | 8 |
Journal | European radiology |
Volume | 28 |
Issue number | 11 |
DOIs | |
Publication status | Published - 1 Nov 2018 |
Keywords
- Image-guided biopsy
- Magnetic resonance imaging
- Operative time
- Prostate cancer
- Tablet computers