High failure rate but promising clinical performance after implantation of a flexible medial meniscus prosthesis at 1‐year follow‐up

Branco S. van Minnen*, Petra J.C. Heesterbeek, Koen C. Defoort, Pieter J. Emans, Ewoud R.A. van Arkel, Thijmen Struik, Liesbeth M. Jutten, Saskia Susan, Sebastiaan A.W. van de Groes, Nico Verdonschot, Tony G. van Tienen

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

11 Downloads (Pure)

Abstract

Purpose
After unsatisfactory results in a first-in-human clinical investigation with an anatomically shaped medial meniscus prosthesis, the prosthesis and its fixation technique were altered. This interim analysis of a prospective single-arm clinical investigation aims to evaluate safety and clinical performance in a first-in-human study with the redesigned meniscus prosthesis system.

Methods
Ten patients suffering from medial postmeniscectomy pain syndrome were treated with the meniscus prosthesis. Patient-reported outcome measures were obtained at baseline and at 6-week, 3-, 6-, and 12-month follow-up. Radiographs and magnetic resonance imaging scans were obtained to evaluate joint degeneration and prosthesis location.

Results
The device alterations resolved the issues occurring with the previous design, but four prostheses were explanted after fixation failure or subluxation of the prosthesis. Five out of six patients who reached 1-year follow-up reported a clinically significant improvement of the knee injury and osteoarthritis outcome score pain subscale. Imaging revealed no adverse effects on joint degeneration.

Conclusion
The failure mechanisms that occurred with the previous design have been resolved, but the new fixation technique introduced new safety issues. Improvement of positioning and fixation techniques are considered essential future adaptations to reduce the risk of failure. The good clinical outcomes reported by the patients reaching 1-year follow-up indicate that the medial meniscus prosthesis is a potential solution for patients suffering from postmeniscectomy pain syndrome.

Level of Evidence
Level II.
Original languageEnglish
Number of pages11
JournalKnee surgery, sports traumatology, arthroscopy
Early online date3 Sept 2024
DOIs
Publication statusE-pub ahead of print/First online - 3 Sept 2024
Externally publishedYes

Fingerprint

Dive into the research topics of 'High failure rate but promising clinical performance after implantation of a flexible medial meniscus prosthesis at 1‐year follow‐up'. Together they form a unique fingerprint.

Cite this