In Vitro Quantification of Gutter Formation and Chimney Graft Compression in Chimney EVAR Stent-Graft Configurations Using Electrocardiography-Gated Computed Tomography

Simon P. Overeem* (Corresponding Author), Esmé J. Donselaar, Jorrit T. Boersen, Erik Groot Jebbink, Cornelis H. Slump, Jean-Paul P.M. de Vries, Michel M.P.J. Reijnen

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

7 Citations (Scopus)
1 Downloads (Pure)

Abstract

Purpose: To assess the dynamic behavior of chimney grafts during the cardiac cycle. Methods: Three chimney endovascular aneurysm repair (EVAR) stent-graft configurations (Endurant and Advanta V12, Endurant and Viabahn, and Endurant and BeGraft) were placed in silicone aneurysm models and subjected to physiologic flow. Electrocardiography (ECG)-gated contrast-enhanced computed tomography was used to visualize geometric changes during the cardiac cycle. Endograft and chimney graft surface, gutter volume, chimney graft angulation over the center lumen line, and the D-ratio (the ratio between the lengths of the major and minor axes) were independently assessed by 2 observers at 10 time points in the cardiac cycle. Results: Both gutter volumes and chimney graft geometry changed significantly during the cardiac cycle in all 3 configurations (p<0.001). Gutters and endoleaks were observed in all configurations. The largest gutter volume (232.8 mm3) and change in volume (20.7 mm3) between systole and diastole were observed in the Endurant-Advanta configuration. These values were 2.7- and 3.0-fold higher, respectively, compared to the Endurant-Viabahn configuration and 1.7- and 1.6-fold higher as observed in the Endurant-BeGraft configuration. The Endurant-Viabahn configuration had the highest D-ratio (right, 1.26–1.35; left, 1.33–1.48), while the Endurant-BeGraft configuration had the lowest (right, 1.11–1.17; left, 1.08–1.15). Assessment of the interobserver variability showed a high correlation (intraclass correlation >0.935) between measurements. Conclusion: Gutter volumes and stent compression are dynamic phenomena that reshape during the cardiac cycle. Compelling differences were observed during the cardiac cycle in all configurations, with the self-expanding (Endurant–Viabahn) chimney EVAR configurations having smaller gutters and less variation in gutter volume during the cardiac cycle yet more stent compression without affecting the chimney graft surface.

Original languageEnglish
Pages (from-to)387-394
Number of pages8
JournalJournal of Endovascular Therapy
Volume25
Issue number3
DOIs
Publication statusPublished - 1 Jun 2018

Keywords

  • UT-Hybrid-D
  • aneurysm diameter
  • aneurysm model
  • cardiac cycle
  • chimney graft
  • dynamic computed tomography
  • ECG-gated computed tomography
  • endograft
  • endoleak
  • endovascular aneurysm repair
  • flow model
  • gutter
  • in vitro study
  • self-expanding stent-graft
  • stent compression
  • stent-graft
  • volume
  • abdominal aortic aneurysm