In vivo B1+ enhancement of calf MRI at 7 T via optimized flexible metasurfaces

Paul S. Jacobs, Neil Wilson, Wyger Brink, Anshuman Swain, Ryan Armbruster, Aniketh Hanumapur, M. Dylan Tisdall, John Detre, Ravi Prakash Reddy Nanga, Mark A. Elliott, Ravinder Reddy*

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

46 Downloads (Pure)

Abstract

Purpose: Ultrahigh field (≥7 T) MRI is at the cutting edge of medical imaging, enabling enhanced spatial and spectral resolution as well as enhanced susceptibility contrast. However, transmit ((Formula presented.)) field inhomogeneity due to standing wave effects caused by the shortened RF wavelengths at 7 T is still a challenge to overcome. Novel hardware methods such as dielectric pads have been shown to improve the (Formula presented.) field inhomogeneity but are currently limited in their corrective effect by the range of high-permittivity materials available and have a fixed shelf life. In this work, an optimized metasurface design is presented that demonstrates in vivo enhancement of the (Formula presented.) field. Methods: A prototype metasurface was optimized by an empirical capacitor sweep and by varying the period size. Phantom temperature experiments were performed to evaluate potential metasurface heating effects during scanning. Lastly, in vivo gradient echo images and (Formula presented.) maps were acquired on five healthy subjects on a 7 T system. Dielectric pads were also used as a comparison throughout the work as a standard comparison. Results: The metasurfaces presented here enhanced the average relative SNR of the gradient echo images by a factor of 2.26 compared to the dielectric pads factor of 1.61. Average (Formula presented.) values reflected a similar enhancement of 27.6% with the metasurfaces present versus 8.9% with the dielectric pads. Conclusion: The results demonstrate that metasurfaces provide superior performance to dielectric padding as shown by (Formula presented.) maps reflecting their direct effects and resulting enhancements in image SNR at 7 T.

Original languageEnglish
Pages (from-to)1277-1289
Number of pages13
JournalMagnetic resonance in medicine
Volume92
Issue number3
Early online date12 Mar 2024
DOIs
Publication statusPublished - Sept 2024

Keywords

  • 2024 OA procedure
  • metasurfaces
  • Ultra-high field
  • B1+ enhancement

Fingerprint

Dive into the research topics of 'In vivo B1+ enhancement of calf MRI at 7 T via optimized flexible metasurfaces'. Together they form a unique fingerprint.

Cite this