Skip to main navigation Skip to search Skip to main content

In Vivo Brain B1+ Inhomogeneity Correction and NOE Image Enhancement at 7 T via Flexible Metasurfaces

  • Paul S Jacobs
  • , Anshuman Swain
  • , Neil E Wilson
  • , Wyger M Brink
  • , Abeer Mathur
  • , Ravi Prakash Reddy Nanga
  • , Mark A Elliott
  • , John A Detre
  • , Ravinder Reddy*
  • *Corresponding author for this work

    Research output: Contribution to journalArticleAcademicpeer-review

    14 Downloads (Pure)

    Abstract

    Nuclear Overhauser effect (NOE) MRI has been used for in vivo brain imaging to assess lipid and protein composition and benefits from 7 T field strengths due to the larger chemical shift dispersion. However, a continuing challenge is signal drop off observed in regions such as the medial temporal lobes due to “standing wave” effects from shorter radiofrequency (RF) wavelengths at ultra-high fields. 2D periodic unit cell metasurfaces have been a promising approach for providing improvements in anatomical imaging but have not yet been evaluated in chemical exchange saturation transfer (CEST)-based sequences. Here, we report the use of metasurfaces for enhancement of NOE imaging as well as for improvement of Lorentzian line fitting of full Z-spectrum data. 3D NOE image data, B1+ maps, and B0 maps were acquired on five healthy volunteers using a 7 T MRI system with and without metasurfaces positioned near the temporal lobes. A frequency offset range of −5 to +5 ppm with additional separate acquisitions of ±20 and ±100 ppm offset images. A five-pool Lorentzian line fitting model was employed to fit and quantitatively compared magnetization transfer (MT), amide proton transfer (APT), amine, and relayed NOE (rNOE) metabolite pools. NOEMTR-weighted contrast maps were also calculated via Z-spectrum asymmetry analysis. The metasurfaces globally enhanced the transmit efficiency within the imaging slab by approximately 9.6% and reduced B1+ inhomogeneity by approximately 16.6% and increased transmit efficiency by 55.8% in the temporal lobes. Amplitude fit maps showed decreases in contrast magnitude ranging from 1 to 16% and changes in image uniformity ranging from a 4.3 decrease to a 34.7% increase, while NOEMTR-weighted contrast maps demonstrated similar changes. The results presented here demonstrate that metasurfaces can enhance CEST-based techniques complementing previously reported benefits in anatomical imaging.

    Original languageEnglish
    Article numbere70027
    JournalNMR in biomedicine
    Volume38
    Issue number5
    DOIs
    Publication statusPublished - May 2025

    Keywords

    • B homogeneity
    • metasurfaces
    • NOE
    • ultra-high field

    Fingerprint

    Dive into the research topics of 'In Vivo Brain B1+ Inhomogeneity Correction and NOE Image Enhancement at 7 T via Flexible Metasurfaces'. Together they form a unique fingerprint.

    Cite this