TY - JOUR
T1 - Intra-operative fiducial-based CT/fluoroscope image registration framework for image-guided robot-assisted joint fracture surgery
AU - Dagnino, Giulio
AU - Georgilas, Ioannis
AU - Morad, Samir
AU - Gibbons, Peter
AU - Tarassoli, Payam
AU - Atkins, Roger
AU - Dogramadzi, Sanja
PY - 2017/8/1
Y1 - 2017/8/1
N2 - Purpose: Joint fractures must be accurately reduced minimising soft tissue damages to avoid negative surgical outcomes. To this regard, we have developed the RAFS surgical system, which allows the percutaneous reduction of intra-articular fractures and provides intra-operative real-time 3D image guidance to the surgeon. Earlier experiments showed the effectiveness of the RAFS system on phantoms, but also key issues which precluded its use in a clinical application. This work proposes a redesign of the RAFS’s navigation system overcoming the earlier version’s issues, aiming to move the RAFS system into a surgical environment. Methods: The navigation system is improved through an image registration framework allowing the intra-operative registration between pre-operative CT images and intra-operative fluoroscopic images of a fractured bone using a custom-made fiducial marker. The objective of the registration is to estimate the relative pose between a bone fragment and an orthopaedic manipulation pin inserted into it intra-operatively. The actual pose of the bone fragment can be updated in real time using an optical tracker, enabling the image guidance. Results: Experiments on phantom and cadavers demonstrated the accuracy and reliability of the registration framework, showing a reduction accuracy (sTRE) of about 0.88±0.2mm (phantom) and 1.15±0.8mm (cadavers). Four distal femur fractures were successfully reduced in cadaveric specimens using the improved navigation system and the RAFS system following the new clinical workflow (reduction error 1.2±0.3mm, 2 ± 1 ∘). Conclusion: Experiments showed the feasibility of the image registration framework. It was successfully integrated into the navigation system, allowing the use of the RAFS system in a realistic surgical application.
AB - Purpose: Joint fractures must be accurately reduced minimising soft tissue damages to avoid negative surgical outcomes. To this regard, we have developed the RAFS surgical system, which allows the percutaneous reduction of intra-articular fractures and provides intra-operative real-time 3D image guidance to the surgeon. Earlier experiments showed the effectiveness of the RAFS system on phantoms, but also key issues which precluded its use in a clinical application. This work proposes a redesign of the RAFS’s navigation system overcoming the earlier version’s issues, aiming to move the RAFS system into a surgical environment. Methods: The navigation system is improved through an image registration framework allowing the intra-operative registration between pre-operative CT images and intra-operative fluoroscopic images of a fractured bone using a custom-made fiducial marker. The objective of the registration is to estimate the relative pose between a bone fragment and an orthopaedic manipulation pin inserted into it intra-operatively. The actual pose of the bone fragment can be updated in real time using an optical tracker, enabling the image guidance. Results: Experiments on phantom and cadavers demonstrated the accuracy and reliability of the registration framework, showing a reduction accuracy (sTRE) of about 0.88±0.2mm (phantom) and 1.15±0.8mm (cadavers). Four distal femur fractures were successfully reduced in cadaveric specimens using the improved navigation system and the RAFS system following the new clinical workflow (reduction error 1.2±0.3mm, 2 ± 1 ∘). Conclusion: Experiments showed the feasibility of the image registration framework. It was successfully integrated into the navigation system, allowing the use of the RAFS system in a realistic surgical application.
KW - CT/fluoroscopy data
KW - Image-guided surgery
KW - Intra-operative registration
KW - Minimally invasive fracture surgery
KW - Robot-assisted surgery
UR - http://www.scopus.com/inward/record.url?scp=85018784616&partnerID=8YFLogxK
U2 - 10.1007/s11548-017-1602-9
DO - 10.1007/s11548-017-1602-9
M3 - Article
C2 - 28474269
AN - SCOPUS:85018784616
SN - 1861-6410
VL - 12
SP - 1383
EP - 1397
JO - International journal of computer assisted radiology and surgery
JF - International journal of computer assisted radiology and surgery
IS - 8
ER -