    Abstract

    This paper reviews the results that have been reported on ISFET based enzyme sensors. The most important improvement that results from the application of ISFETs instead of glass membrane electrodes is in the method of fabrication. Problems with regard to the pH dependence of the response and the dynamic range as well as the influence of the sample buffer capacity have not been solved. As a possible solution we introduce a coulometric system that compensates for the analyte buffer capacity. If the pH in the immobilized enzyme layer is thus controlled, the resulting pH-static enzyme sensor has an output that is independent of the sample pH and buffer capacity and has an expanded linear range.
    Original languageEnglish
    Pages (from-to)161-186
    JournalBiosensors (UK)
    Volume3
    Issue number3
    DOIs
    Publication statusPublished - 1987

    Keywords

    • pH-static enzyme sensor
    • Buffer dependence
    • Coulometry
    • Urea
    • IC technology
    • ISFET
    • Enzyme

    Fingerprint Dive into the research topics of 'ISFET based enzyme sensors'. Together they form a unique fingerprint.