N2 - Objective: This study evaluated the long term outcomes of endovascular aneurysm repair using the Gore Excluder Low Permeability (LP) endoprosthesis across high volume Dutch hospitals. Methods: A retrospective analysis was conducted of patients treated with the Excluder LP for infrarenal abdominal aortic aneurysm (AAA) in four hospitals between 2004 and 2017. Primary outcomes were overall survival, freedom from re-interventions (overall, inside and outside instructions for use, IFU), and AAA sac dynamics: growth (> 5 mm), stabilisation, and regression (< 5 mm). Secondary outcomes were technical success (device deployment), procedural parameters, and re-interventions. Follow up visits were extracted from patient files, with imaging assessed for complications and AAA diameter. Results: Five hundred and fourteen patients were enrolled, with a median (IQR) follow up of 5.0 (2.9, 6.9) years. Survival rates were 94.0% at one year, 73.0% at five years, and 37.0% at 10 years, with freedom from re-interventions of 89.0%, 79.0%, and 71.0%, respectively. 37.9% were treated outside IFU, leading to significantly more re-interventions over 10 years compared with those treated inside IFU (36.0% vs. 25.0%, respectively; p = .044). The aneurysm sac regressed by 53.5% at one year, 65.8% at five years, and 77.8% at 10 years, and grew by 9.8%, 14.3%, and 22.2%, respectively. Patients with one year sac growth had significantly worse survival (p = .047). Seven patients (1.4%) had a ruptured aneurysm during follow up. Over 15 years, type 1a endoleak occurred in 5.3%, type 1b in 3.1%, type 3 in 1.9%, type 4 in 0.2%, and type 2 in 35.6% of patients. Conclusion: This multicentre study of real world endovascular aneurysm repair data using the Gore Excluder LP endoprosthesis demonstrated robust long term survival and re-intervention rates, despite 37.9% of patients being treated outside IFU, with type 4 endoleak being rare. Treatment outside IFU significantly increased re-intervention rates and one year sac growth was associated with statistically significantly worse survival.
AB - Objective: This study evaluated the long term outcomes of endovascular aneurysm repair using the Gore Excluder Low Permeability (LP) endoprosthesis across high volume Dutch hospitals. Methods: A retrospective analysis was conducted of patients treated with the Excluder LP for infrarenal abdominal aortic aneurysm (AAA) in four hospitals between 2004 and 2017. Primary outcomes were overall survival, freedom from re-interventions (overall, inside and outside instructions for use, IFU), and AAA sac dynamics: growth (> 5 mm), stabilisation, and regression (< 5 mm). Secondary outcomes were technical success (device deployment), procedural parameters, and re-interventions. Follow up visits were extracted from patient files, with imaging assessed for complications and AAA diameter. Results: Five hundred and fourteen patients were enrolled, with a median (IQR) follow up of 5.0 (2.9, 6.9) years. Survival rates were 94.0% at one year, 73.0% at five years, and 37.0% at 10 years, with freedom from re-interventions of 89.0%, 79.0%, and 71.0%, respectively. 37.9% were treated outside IFU, leading to significantly more re-interventions over 10 years compared with those treated inside IFU (36.0% vs. 25.0%, respectively; p = .044). The aneurysm sac regressed by 53.5% at one year, 65.8% at five years, and 77.8% at 10 years, and grew by 9.8%, 14.3%, and 22.2%, respectively. Patients with one year sac growth had significantly worse survival (p = .047). Seven patients (1.4%) had a ruptured aneurysm during follow up. Over 15 years, type 1a endoleak occurred in 5.3%, type 1b in 3.1%, type 3 in 1.9%, type 4 in 0.2%, and type 2 in 35.6% of patients. Conclusion: This multicentre study of real world endovascular aneurysm repair data using the Gore Excluder LP endoprosthesis demonstrated robust long term survival and re-intervention rates, despite 37.9% of patients being treated outside IFU, with type 4 endoleak being rare. Treatment outside IFU significantly increased re-intervention rates and one year sac growth was associated with statistically significantly worse survival.
KW - n/a OA procedure
KW - Endoleak
KW - Endovascular aneurysm repair
KW - Gore Excluder
KW - Long term survival
KW - Low permeability
KW - Abdominal aortic aneurysm
UR - http://www.scopus.com/inward/record.url?scp=85189957443&partnerID=8YFLogxK
U2 - 10.1016/j.ejvs.2024.03.028
DO - 10.1016/j.ejvs.2024.03.028
M3 - Article
C2 - 38527519
AN - SCOPUS:85189957443
SN - 1078-5884
VL - 68
SP - 18
EP - 27
JO - European journal of vascular and endovascular surgery
JF - European journal of vascular and endovascular surgery
IS - 1
ER -