TY - JOUR
T1 - Mechanical implications of interfacial defects between femoral hip implants and cement
T2 - a finite element analysis of interfacial gaps and interfacial porosity
AU - Scheerlinck, T.
AU - Broos, J.
AU - Janssen, D.
AU - Verdonschot, N.
PY - 2008
Y1 - 2008
N2 - Two types of defect between femoral hip implants and cement have been identified. Interfacial porosity arises from cement shrinkage during curing and presents as pores randomly located along the stem. Interfacial gaps are much larger stem-cement separations caused by air introduced during stem insertion. To investigate the mechanical consequences of both types of defect, a finite element analysis model was created on the basis of a computed tomography image of a Charnley-Kerboul stem, and alternating torsional and transverse loads were applied. The propagation of fatigue cracks within the cement and the rotational stability of the stem were assessed in models simulating increasing amounts of interfacial gaps and pores. Anterior gaps covering at least 30 per cent of the implant surface promoted cement cracks and destabilized the stem. Anterolateral gaps were less destabilizing, but had more potential to promote cracks. In both cases, cracks occurred mainly outside gap regions, in areas where the stem contacted the cement during cyclic loading. Although random interfacial pores did not destabilize the implant, they acted as crack initiators even at low fractions (10 per cent). In conclusion, random interfacial pores were more harmful for the cement mantle integrity than were larger regions of interfacial gaps, although gaps were more detrimental for the rotational stability of the stem.
AB - Two types of defect between femoral hip implants and cement have been identified. Interfacial porosity arises from cement shrinkage during curing and presents as pores randomly located along the stem. Interfacial gaps are much larger stem-cement separations caused by air introduced during stem insertion. To investigate the mechanical consequences of both types of defect, a finite element analysis model was created on the basis of a computed tomography image of a Charnley-Kerboul stem, and alternating torsional and transverse loads were applied. The propagation of fatigue cracks within the cement and the rotational stability of the stem were assessed in models simulating increasing amounts of interfacial gaps and pores. Anterior gaps covering at least 30 per cent of the implant surface promoted cement cracks and destabilized the stem. Anterolateral gaps were less destabilizing, but had more potential to promote cracks. In both cases, cracks occurred mainly outside gap regions, in areas where the stem contacted the cement during cyclic loading. Although random interfacial pores did not destabilize the implant, they acted as crack initiators even at low fractions (10 per cent). In conclusion, random interfacial pores were more harmful for the cement mantle integrity than were larger regions of interfacial gaps, although gaps were more detrimental for the rotational stability of the stem.
U2 - 10.1243/09544119JEIM362
DO - 10.1243/09544119JEIM362
M3 - Article
SN - 0954-4119
VL - 222
SP - 1037
EP - 1047
JO - Proceedings of the Institution of Mechanical Engineers. Part H: Journal of engineering in medicine
JF - Proceedings of the Institution of Mechanical Engineers. Part H: Journal of engineering in medicine