Perception threshold and electrode position for spinal cord stimulation

Jiping He, Giancarlo Barolat, Jan Holsheimer, Johannes J. Struijk

    Research output: Contribution to journalArticleAcademicpeer-review

    66 Citations (Scopus)
    116 Downloads (Pure)

    Abstract

    The perception threshold for epidural spinal cord stimulation in chronic pain management was analyzed on 3923 testing data obtained from 136 implanted patients. The initial areas of paresthesiae due to stimulation were recorded and reported as the stimulation map according to the location of electrodes. Measurement of dorsal thickness of the cerebrospinal fluid (CSF) layer was obtained from 26 subjects using magnetic resonance imaging (MRI). The results indicate that the perception threshold is a function of the spinal level of the implanted electrodes, of the mediolateral position in the spinal canal and the contact separation of electrode. Differences in perception threshold at various vertebral levels are mainly due to varying depths of the dorsal CSF layer. The medially placed electrodes caudal to the mid-cervical levels have a higher perception threshold than more laterally placed ones. The electrodes at high and mid-cervical levels, however, have a smaller perception threshold if placed medially. The information obtained from this investigation has important implications for the design of a new-generation stimulation system and clinical application to maximize the longevity of the power source.
    Original languageEnglish
    Pages (from-to)55-63
    JournalPain
    Volume59
    Issue number1
    DOIs
    Publication statusPublished - 1994

    Keywords

    • Perception threshold
    • spinal cord stimulation (SCS)
    • Electrode combination
    • Cerebrospinal fluid
    • Mediolatera1 position
    • Pain
    • Battery life

    Fingerprint Dive into the research topics of 'Perception threshold and electrode position for spinal cord stimulation'. Together they form a unique fingerprint.

    Cite this