Abstract
A microchip CE-based method for the quantification of the thiols mercaptoethanoic acid (MAA) and 2-mercaptopropionic acid (2-MPA) in depilatory cream and cold wave lotions was developed. The thiols were first derivatized with the fluorogenic reagent ammonium-7-fluorobenzo-2-oxa-1,3-diazole-4-sulfonate (SBD-F). The derivatives were separated within only 20 s by microchip CE and detected by their fluorescence. Conventional CE with diode array detection and LC with fluorescence detection were used for validation. The internal standard 3-mercaptopropionic acid (3-MPA) provided RSDs of multiple injections of only 4% or less for the MCE approach. LOD is 2 µM, LOQ 6 µM, and the linear range comprises nearly three decades of concentration starting at the LOQ.
Original language | Undefined |
---|---|
Pages (from-to) | 1154-1160 |
Journal | Electrophoresis |
Volume | 28 |
Issue number | 7 |
DOIs | |
Publication status | Published - 2007 |
Keywords
- Quantitative analysis
- Microchip CE
- Derivatization
- Fluorescence detection
- IR-72255
- Thiols
- METIS-246554