Scanning electron microscopic assessment of the biodegradable coating on expanded biolimus-eluting stents

Mounir W.Z. Basalus, K. Gert van Houwelingen, Marc Ankoné, Frits H.A.F. de Man, Clemens von Birgelen

    Research output: Contribution to journalArticleAcademicpeer-review

    244 Downloads (Pure)

    Abstract

    AIMS: Biodegradable coatings on drug-eluting stents (DES) may help to avoid adverse long-term effects of DES such as late/very late stent thrombosis which is partly attributed to durable polymers. As the post-expansion morphology of biodegradable coatings is greatly unknown, we investigated the polylactic acid coating on biolimus-eluting BioMatrix stents.

    METHODS AND RESULTS: Scanning electron microscopy (SEM) was used to carefully examine five 3.5mm stents following expansion at 6-14atm (maximum 7% overstretch). SEM examination demonstrated only mild cracks of the coating after stent expansion at 6 atm. An increase in expansion pressure, associated with mild stent overstretch, resulted in more severe cracks. Lifting of the coating together with few sites of partial detachment of fragments was noticed after stent expansion in water at 14atm; these irregularities further increased after aggressive oversized partial postdilatation with a 5.0mm non-compliant balloon with additional secondary cracks.

    CONCLUSIONS: SEM assessment suggests a relatively low elasticity of the biodegradable coating on BioMatrix stents. At nominal pressure, stents showed predominantly mild cracks of the coating, while cracks increased after slight overstretch. Aggressive overexpansion of the stent, such as sometimes required in left main bifurcation stenting, worsened cracks and led to some detachment of fragments of the coating in vitro.
    Original languageEnglish
    Pages (from-to)505-510
    Number of pages6
    JournalEuroIntervention
    Volume5
    Issue number4
    DOIs
    Publication statusPublished - 2009

    Keywords

    • METIS-259910
    • Biodegradable polymer coating
    • Scanning Electron Microscopy (SEM)
    • Drug-eluting stent(s)
    • Percutaneous Coronary Intervention (PCI)
    • Coronary artery disease
    • BioMatrix

    Fingerprint

    Dive into the research topics of 'Scanning electron microscopic assessment of the biodegradable coating on expanded biolimus-eluting stents'. Together they form a unique fingerprint.

    Cite this