Abstract
We demonstrate a pinched flow tangential filtration method to sort spermatozoa from larger particles with a spermatozoa collection efficiency of 94±2% and a separation efficiency of 100%. In conventional pinched flow fractionation (PFF), an observed tumbling-like rotation of spermatozoa complicates their separation from other articles. The new design circumvents this effect.
Original language | Undefined |
---|---|
Title of host publication | 20th International Conference on Miniaturized Systems for Chemistry and LifeSciences, MicroTAS 2016 |
Publisher | The Chemical and Biological Microsystems Society |
Pages | 1587-1588 |
Number of pages | 2 |
ISBN (Print) | 978-0-9798064-9-0 |
Publication status | Published - 16 Oct 2016 |
Event | 20th International Conference on Miniaturized Systems for Chemistry and LifeSciences, µTAS 2016 - Convention Center Dublin, Dublin, Ireland Duration: 9 Oct 2016 → 13 Oct 2016 Conference number: 20 http://www.microtas2016.org/ |
Publication series
Name | International conference on Miniaturized Systems for Chemistry and LifeSciences (MicroTAS) |
---|---|
Publisher | The Chemical and Biological Microsystems Society |
Number | -- |
ISSN (Print) | 1556-5904 |
Conference
Conference | 20th International Conference on Miniaturized Systems for Chemistry and LifeSciences, µTAS 2016 |
---|---|
Abbreviated title | MicroTAS 2016 |
Country/Territory | Ireland |
City | Dublin |
Period | 9/10/16 → 13/10/16 |
Internet address |
Keywords
- EWI-27516
- IR-102626
- METIS-319513