Simple test for electrophysiologically tolerable parameters of artificial stimulation

Piet Bergveld

Research output: Contribution to journalArticleAcademic

6 Citations (Scopus)
50 Downloads (Pure)

Abstract

Electrical stimulation of irritable tissue by a current source not only provides well defined stimulation but gives also the possibility to test whether the pulse amplitude, duration and rate are tolerable or not in order to prevent undesirable electrode reactions. This test can be very easily performed by recording the voltage developed between the stimulating electrodes as a function of time. This voltage must never reach constant value after a period of time.
Original languageUndefined
Pages (from-to)479-482
JournalMedical & biological engineering & computing
Volume14
Issue number5
DOIs
Publication statusPublished - 1976

Keywords

  • Artificial stimulation
  • IR-85594
  • Stimulation electrodes
  • Stimulation parameters

Cite this

@article{7e46c04a3d304677bde40394517aa89b,
title = "Simple test for electrophysiologically tolerable parameters of artificial stimulation",
abstract = "Electrical stimulation of irritable tissue by a current source not only provides well defined stimulation but gives also the possibility to test whether the pulse amplitude, duration and rate are tolerable or not in order to prevent undesirable electrode reactions. This test can be very easily performed by recording the voltage developed between the stimulating electrodes as a function of time. This voltage must never reach constant value after a period of time.",
keywords = "Artificial stimulation, IR-85594, Stimulation electrodes, Stimulation parameters",
author = "Piet Bergveld",
year = "1976",
doi = "10.1007/BF02478042",
language = "Undefined",
volume = "14",
pages = "479--482",
journal = "Medical & biological engineering & computing",
issn = "0140-0118",
publisher = "Springer",
number = "5",

}

Simple test for electrophysiologically tolerable parameters of artificial stimulation. / Bergveld, Piet.

In: Medical & biological engineering & computing, Vol. 14, No. 5, 1976, p. 479-482.

Research output: Contribution to journalArticleAcademic

TY - JOUR

T1 - Simple test for electrophysiologically tolerable parameters of artificial stimulation

AU - Bergveld, Piet

PY - 1976

Y1 - 1976

N2 - Electrical stimulation of irritable tissue by a current source not only provides well defined stimulation but gives also the possibility to test whether the pulse amplitude, duration and rate are tolerable or not in order to prevent undesirable electrode reactions. This test can be very easily performed by recording the voltage developed between the stimulating electrodes as a function of time. This voltage must never reach constant value after a period of time.

AB - Electrical stimulation of irritable tissue by a current source not only provides well defined stimulation but gives also the possibility to test whether the pulse amplitude, duration and rate are tolerable or not in order to prevent undesirable electrode reactions. This test can be very easily performed by recording the voltage developed between the stimulating electrodes as a function of time. This voltage must never reach constant value after a period of time.

KW - Artificial stimulation

KW - IR-85594

KW - Stimulation electrodes

KW - Stimulation parameters

U2 - 10.1007/BF02478042

DO - 10.1007/BF02478042

M3 - Article

VL - 14

SP - 479

EP - 482

JO - Medical & biological engineering & computing

JF - Medical & biological engineering & computing

SN - 0140-0118

IS - 5

ER -