Temperature dependence of the effective sensing area of high-Τc dc SQUIDs

    Research output: Contribution to journal › Article › Academic › peer-review

    4 Downloads (Pure)

    Abstract

    The effective sensing area of a high-Tc dc SQUID depends on temperature. As a consequence, fluctuations in the operating temperature result in apparent magnetic field noise if the SQUID is placed in a background magnetic field. An analysis of this effect for two SQUID types, the square-washer 'Ketchen' type and the inductively shunted type, is performed. For magnetocardiography, the temperature fluctuations (peak to peak) of the latter SQUID type should be below w0.3 mK at 77 K, and below 2 mK at 55 K, with an earth's field suppression of 40 dB. For the square-washer SQUID the requirements are about 8 times less stringent.
    Original languageEnglish
    Pages (from-to)512-515
    Number of pages4
    JournalSuperconductor science and technology
    Volume10
    Issue number7
    DOIs
    Publication statusPublished - 1997