Abstract
The resonance frequencies of individual SonoVue/spl trade/ contrast agent bubbles were measured optically by recording the radius-time curves of a single microbubble at 24 different frequencies. For these experiments, the Brandaris 128 fast framing camera was operated in a special segmented mode. The resonance frequencies found for SonoVue/spl trade/ microbubbles are in good agreement with the modified Herring model for coated bubbles indicating that the shell only slightly affects the resonance frequency of this class of contrast bubbles.
Original language | Undefined |
---|---|
Pages | 343-345 |
DOIs | |
Publication status | Published - 2004 |
Event | 2004 IEEE Ultrasonics Symposium - Montreal, Canada Duration: 23 Aug 2004 → 27 Aug 2004 |
Conference
Conference | 2004 IEEE Ultrasonics Symposium |
---|---|
Country/Territory | Canada |
City | Montreal |
Period | 23/08/04 → 27/08/04 |
Keywords
- IR-55805