Thermal flow measurements at Gr/Re2 ≫ 1 by silicon anemometry

Michael J.A.M. Van Putten, Maurice H.P.M. Van Putten, Anton F.P. Van Putten

Research output: Contribution to journalArticleAcademicpeer-review

15 Citations (Scopus)

Abstract

We present the first characteristics of air flow measurements in the mixed convective region with strong buoyancy effects (Gr/Re2 ≫ 1), using a thermal silicon double-Wheatstone-bridge vector sensor. The Alternating Direction Method is applied to eliminate additive drift influences. The upper limit on the mixed convective region is experimentally defined to be given by Gr/Re2 = [0.3-0.8]. Flow velocity measurements of down to 1 mm/s (Reynolds ≈ 0.3) are realized with an accuracy better than 98%.

Original languageEnglish
Pages (from-to)724-729
Number of pages6
JournalIEEE transactions on instrumentation and measurement
Volume48
Issue number3
DOIs
Publication statusPublished - 1999
Externally publishedYes

Keywords

  • n/a OA procedure

Fingerprint

Dive into the research topics of 'Thermal flow measurements at Gr/Re2 ≫ 1 by silicon anemometry'. Together they form a unique fingerprint.