Thin Composite-Wire-Strut Zotarolimus-Eluting Stents Versus Ultrathin-Strut Sirolimus-Eluting Stents in BIONYX at 2 Years

  • Rosaly A. Buiten
  • , Eline H. Ploumen
  • , Paolo Zocca
  • , Carine J.M. Doggen
  • , Gillian A.J. Jessurun
  • , Carl E. Schotborgh
  • , Ariel Roguin
  • , Peter W. Danse
  • , Edouard Benit
  • , Adel Aminian
  • , K. Gert van Houwelingen
  • , Alexander R. Schramm
  • , Martin G. Stoel
  • , Samer Somi
  • , Marc Hartmann
  • , Gerard C.M. Linssen
  • , Clemens von Birgelen*
  • *Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

30 Citations (Scopus)
139 Downloads (Pure)

Abstract

Objectives: The aim of this study was to assess 2-year safety and efficacy of the current-generation thin composite-wire-strut durable-polymer Resolute Onyx zotarolimus-eluting stent (ZES), compared with the ultrathin-strut biodegradable-polymer Orsiro sirolimus-eluting stent (SES) in all-comers and a pre-specified small-vessel subgroup analysis. Background: The Resolute Onyx ZES is widely used in clinical practice, but no follow-up data beyond 1 year have been published. The randomized BIONYX (Bioresorbable Polymer-Coated Orsiro Versus Durable Polymer-Coated Resolute Onyx Stents) trial (NCT02508714) established the noninferiority of ZES versus SES regarding target vessel failure (TVF) rates. Methods: A total of 2,488 all-comer patients were treated at 7 coronary intervention centers in Belgium, Israel, and the Netherlands. The main endpoint, TVF, was a composite of safety (cardiac death or target vessel–related myocardial infarction) and efficacy (clinically indicated target vessel revascularization). Two-year follow-up data were analyzed using Kaplan-Meier methods. Results: Two-year follow-up data were available for 2,460 of 2,488 patients (98.9%). TVF occurred in 93 of 1,243 patients (7.6%) assigned to ZES versus 87 of 1,245 patients (7.1%) assigned to SES (log-rank p = 0.66). There was no significant between-stent difference in individual components of this endpoint. The incidence of definite-or-probable stent thrombosis was low for both treatment arms (0.4% vs. 1.1%; log-rank p = 0.057). In patients stented in small vessels, there was no between-stent difference (TVF 8.2% vs. 8.7% [log-rank p = 0.75], target lesion revascularization 4.0% vs. 4.4% [log-rank p = 0.77]). Conclusions: At 2-year follow-up, the novel thin composite-wire-strut durable-polymer Resolute Onyx ZES showed in all-comers similar safety and efficacy compared with the ultrathin cobalt-chromium-strut biodegradable-polymer Orsiro SES. The analysis of patients who were treated in small vessels also suggested no advantage for either stent.

Original languageEnglish
Pages (from-to)1100-1109
Number of pages10
JournalJACC: Cardiovascular Interventions
Volume13
Issue number9
Early online date4 May 2020
DOIs
Publication statusPublished - 11 May 2020

Keywords

  • clinical trial
  • drug-eluting stents
  • percutaneous coronary intervention