T2-weighted 3D fMRI using S2-SSFP at 7 tesla

Markus Barth*, Heiko Meyer, Stephan A.R. Kannengiesser, Jonathan R. Polimeni, Lawrence L. Wald, David G. Norris

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

34 Citations (Scopus)

Abstract

In this study, the sensitivity of the S2-steady-state free precession (SSFP) signal for functional MRI at 7 T was investigated. In order to achieve the necessary temporal resolution, a three-dimensional acquisition scheme with acceleration along two spatial axes was employed. Activation maps based on S2-steady-state free precession data showed similar spatial localization of activation and sensitivity as spin-echo echo-planar imaging (SE-EPI), but data can be acquired with substantially lower power deposition. The functional sensitivity estimated by the average z-values was not significantly different for SE-EPI compared to the S2-signal but was slightly lower for the S2-signal (6.74 ± 0.32 for the TR = 15 ms protocol and 7.51 ± 0.78 for the TR = 27 ms protocol) compared to SE-EPI (7.49 ± 1.44 and 8.05 ± 1.67) using the same activated voxels, respectively. The relative signal changes in these voxels upon activation were slightly lower for SE-EPI (2.37% ± 0.18%) compared to the TR 5 15 ms S2-SSFP protocol (2.75% ± 0.53%) and significantly lower than the TR = 27 ms protocol (5.38% ± 1.28%), in line with simulations results. The large relative signal change for the long TR SSFP protocol can be explained by contributions from multiple coherence pathways and the low intrinsic intensity of the S2 signal. In conclusion, whole-brain T2-weighted functional MRI with negligible image distortion at 7 T is feasible using the S2-SSFP sequence and partially parallel imaging.

Original languageEnglish
Pages (from-to)1015-1020
Number of pages6
JournalMagnetic resonance in medicine
Volume63
Issue number4
DOIs
Publication statusPublished - May 2010
Externally publishedYes

Keywords

  • 7 tesla
  • BOLD
  • Brain
  • Neuroimaging
  • Steady-state free procession
  • n/a OA procedure

Fingerprint

Dive into the research topics of 'T2-weighted 3D fMRI using S2-SSFP at 7 tesla'. Together they form a unique fingerprint.