TY - JOUR
T1 - Use of Ventricular Assist Device in Univentricular Physiology
T2 - The Role of Lumped Parameter Models
AU - Di Molfetta, Arianna
AU - Ferrari, Gianfranco
AU - Filippelli, Sergio
AU - Fresiello, Libera
AU - Iacobelli, Roberta
AU - Gagliardi, Maria G.
AU - Amodeo, Antonio
N1 - Publisher Copyright:
© 2016 Wiley Periodicals, Inc. and International Center for Artificial Organs and Transplantation.
PY - 2016/5/1
Y1 - 2016/5/1
N2 - Failing single-ventricle (SV) patients might benefit from ventricular assist devices (VADs) as a bridge to heart transplantation. Considering the complex physiopathology of SV patients and the lack of established experience, the aim of this work was to realize and test a lumped parameter model of the cardiovascular system, able to simulate SV hemodynamics and VAD implantation effects. Data of 30 SV patients (10 Norwood, 10 Glenn, and 10 Fontan) were retrospectively collected and used to simulate patients' baseline. Then, the effects of VAD implantation were simulated. Additionally, both the effects of ventricular assistance and cavopulmonary assistance were simulated in different pathologic conditions on Fontan patients, including systolic dysfunction, diastolic dysfunction, and pulmonary vascular resistance increment. The model can reproduce patients' baseline well. Simulation results suggest that the implantation of VAD: (i) increases the cardiac output (CO) in all the three palliation conditions (Norwood 77.2%, Glenn 38.6%, and Fontan 17.2%); (ii) decreases the SV external work (SVEW) (Norwood 55%, Glenn 35.6%, and Fontan 41%); (iii) increases the mean pulmonary arterial pressure (Pap) (Norwood 39.7%, Glenn 12.1%, and Fontan 3%). In Fontan circulation, with systolic dysfunction, the left VAD (LVAD) increases CO (35%), while the right VAD (RVAD) determines a decrement of inferior vena cava pressure (Pvci) (39%) with 34% increment of CO. With diastolic dysfunction, the LVAD increases CO (42%) and the RVAD decreases the Pvci. With pulmonary vascular resistance increment, the RVAD allows the highest CO (50%) increment with the highest decrement of Pvci (53%). The single ventricular external work (SVEW) increases (decreases) increasing the VAD speed in cavopulmonary (ventricular) assistance. Numeric models could be helpful in this challenging and innovative field to support patients and VAD selection to optimize the clinical outcome and personalize the therapy.
AB - Failing single-ventricle (SV) patients might benefit from ventricular assist devices (VADs) as a bridge to heart transplantation. Considering the complex physiopathology of SV patients and the lack of established experience, the aim of this work was to realize and test a lumped parameter model of the cardiovascular system, able to simulate SV hemodynamics and VAD implantation effects. Data of 30 SV patients (10 Norwood, 10 Glenn, and 10 Fontan) were retrospectively collected and used to simulate patients' baseline. Then, the effects of VAD implantation were simulated. Additionally, both the effects of ventricular assistance and cavopulmonary assistance were simulated in different pathologic conditions on Fontan patients, including systolic dysfunction, diastolic dysfunction, and pulmonary vascular resistance increment. The model can reproduce patients' baseline well. Simulation results suggest that the implantation of VAD: (i) increases the cardiac output (CO) in all the three palliation conditions (Norwood 77.2%, Glenn 38.6%, and Fontan 17.2%); (ii) decreases the SV external work (SVEW) (Norwood 55%, Glenn 35.6%, and Fontan 41%); (iii) increases the mean pulmonary arterial pressure (Pap) (Norwood 39.7%, Glenn 12.1%, and Fontan 3%). In Fontan circulation, with systolic dysfunction, the left VAD (LVAD) increases CO (35%), while the right VAD (RVAD) determines a decrement of inferior vena cava pressure (Pvci) (39%) with 34% increment of CO. With diastolic dysfunction, the LVAD increases CO (42%) and the RVAD decreases the Pvci. With pulmonary vascular resistance increment, the RVAD allows the highest CO (50%) increment with the highest decrement of Pvci (53%). The single ventricular external work (SVEW) increases (decreases) increasing the VAD speed in cavopulmonary (ventricular) assistance. Numeric models could be helpful in this challenging and innovative field to support patients and VAD selection to optimize the clinical outcome and personalize the therapy.
KW - Congenital heart disease
KW - Lumped parameter model
KW - Ventricular assist device
UR - http://www.scopus.com/inward/record.url?scp=84949653973&partnerID=8YFLogxK
U2 - 10.1111/aor.12583
DO - 10.1111/aor.12583
M3 - Article
C2 - 26494529
AN - SCOPUS:84949653973
SN - 0160-564X
VL - 40
SP - 444
EP - 453
JO - Artificial organs
JF - Artificial organs
IS - 5
ER -