Validity and reliability of Eforto®, a system to (self-)monitor grip strength and muscle fatigability in older persons

Liza De Dobbeleer, Myrthe Manouk Swart, Merle Anne Joëlle Geerds, Remco Johan Baggen, Anne-Jet Sophie Jansen, Rudi Tielemans, Hugo Silva, Siddhartha Lieten, Kurt Barbé, Geeske Peeters, Miriam Marie Rosé Vollenbroek-Hutten, René Johannes Franciscus Melis*, Ivan Bautmans*

*Corresponding author for this work

Research output: Contribution to journalArticleAcademicpeer-review

6 Citations (Scopus)
15 Downloads (Pure)

Abstract

Introduction: We developed Eforto®, an innovative system for (self-)monitoring of grip strength (GS) and muscle fatigability (Fatigue Resistance (FR = time until GS decreased to 50% of maximum during sustained contraction) and grip work (GW = area under the strength-time curve)). The Eforto® system consists of a rubber bulb that is wirelessly connected to a smartphone-based application, and a telemonitoring platform. The aim was to evaluate the validity and reliability of Eforto® to measure muscle fatigability. Methods: Community-dwelling older persons (n = 61), geriatric inpatients (n = 26) and hip fracture patients (n = 25) were evaluated for GS and muscle fatigability. In community dwellers fatigability was tested twice in the clinic (once with Eforto®, once with Martin Vigorimeter (MV), standard analog handgrip system) and for six consecutive days as a self-assessment at home with Eforto®. In hospitalized participants, fatigability was tested twice using Eforto®, once by a researcher and once by a health professional. Results: Criterion validity was supported by good to excellent correlations between Eforto® and MV for GS (r = 0.95) and muscle fatigability (FR r = 0.81 and GW r = 0.73), and no significant differences in measurements between both systems. Inter-rater and intra-rater reliability for GW were moderate to excellent (intra-class correlation: 0.59–0.94). The standard error of measurement for GW was small for geriatric inpatients and hip fracture patients (224.5 and 386.5 kPa*s) and higher for community-dwellers (661.5 kPa*s). Discussion/conclusion: We established the criterion validity and reliability of Eforto® in older community-dwelling persons and hospitalized patients, supporting the implementation of Eforto® for (self-)monitoring of muscle fatigability.

Original languageEnglish
Pages (from-to)835-845
Number of pages11
JournalAging clinical and experimental research
Volume35
Issue number4
DOIs
Publication statusPublished - Apr 2023
Externally publishedYes

Keywords

  • Frailty
  • Intrinsic capacity
  • Muscle fatigability
  • Prevention
  • Self-monitoring