Vibration isolation for Coriolis Mass-Flow meters

Bert van de Ridder

Research output: ThesisPhD Thesis - Research UT, graduation UT

640 Downloads (Pure)

Abstract

A Coriolis Mass-Flow Meter (CMFM) is an active device based on the Coriolis force principle for direct mass-flow measurements, with high accuracy, range-ability and repeatability. The working principle of a CMFM is as follows: a fluid conveying tube is actuated to oscillate at a low amplitude. A fluid-flow in the vibrating tube induces Coriolis forces, which are proportional to the mass-flow, and affect the tube motion. Measuring the tube displacement in such a way that the change of its mode shape is determined, allows calculating the mass-flow. For low flows (< 1 kg/h), the Coriolis force induced motion is relatively small compared to motions induced by external vibrations, thus CMFMs designed to be sensitive to low flows are rather sensitive to external vibrations. Mainly external vibrations around the meter’s drive and Coriolis frequencies result in a measurement error. A model-based, quantitative estimation of the expected mass-flow error in response to external vibrations is obtained. The sensitivity to external vibrations is reduced, using passive and active vibration isolation solutions. Passive vibration isolation consists of a mass-spring-damper system between the floor and the measurement device. The performance is insufficient, because the suspension frequency is limited by the maximum stress in the connection tubes and a maximum allowable sag due to gravity. Therefore the passively suspended stage is extended with voice-coil actuators and absolute motion sensors to apply active vibration isolation control. Increased narrow-band vibration attenuation is achieved using loop shaping feedback control. Alternatively, an innovative feed-forward strategy is used, where by the parameters, which are tuned using an adaptive FxLMS technique, are based on the damping and stiffness properties of the stage. Based on the theoretic models, a proof of principle mechanism is realised for validation, whereby attenuation of more than 40 dB is achieved. The performance is limited by sensor noise and for the feedback solution also by the destabilising effect of higher order dynamics. The thesis presents a comprehensive analysis on how vibration isolation can be achieved. Narrow-band active vibration isolation control is successfully implemented in a design of a CMFM, for the first time.
Original languageEnglish
Awarding Institution
  • University of Twente
Supervisors/Advisors
  • de Boer, Andries , Supervisor
  • van Dijk, Johannes , Advisor
  • Hakvoort, Wouter Bernardus Johannes, Advisor
  • de Boer, A, Supervisor
  • van Dijk, J., Advisor
  • Hakvoort, W.B.J., Advisor
Award date24 Jun 2015
Place of PublicationEnschede
Publisher
Print ISBNs978-90-365-3874-9
DOIs
Publication statusPublished - 24 Jun 2015

Fingerprint

mass flow
isolation
vibration
tubes
narrowband
attenuation
theses
flow measurement
modal response
sensors
dampers
feedback control
fluid flow
stiffness
coils
actuators
damping
gravitation

Keywords

  • METIS-310724
  • IR-96105

Cite this

van de Ridder, B. (2015). Vibration isolation for Coriolis Mass-Flow meters. Enschede: University of Twente. https://doi.org/10.3990/1.9789036538749
van de Ridder, Bert. / Vibration isolation for Coriolis Mass-Flow meters. Enschede : University of Twente, 2015. 166 p.
@phdthesis{a7842af50491414a893ce0fd8fc73ab2,
title = "Vibration isolation for Coriolis Mass-Flow meters",
abstract = "A Coriolis Mass-Flow Meter (CMFM) is an active device based on the Coriolis force principle for direct mass-flow measurements, with high accuracy, range-ability and repeatability. The working principle of a CMFM is as follows: a fluid conveying tube is actuated to oscillate at a low amplitude. A fluid-flow in the vibrating tube induces Coriolis forces, which are proportional to the mass-flow, and affect the tube motion. Measuring the tube displacement in such a way that the change of its mode shape is determined, allows calculating the mass-flow. For low flows (< 1 kg/h), the Coriolis force induced motion is relatively small compared to motions induced by external vibrations, thus CMFMs designed to be sensitive to low flows are rather sensitive to external vibrations. Mainly external vibrations around the meter’s drive and Coriolis frequencies result in a measurement error. A model-based, quantitative estimation of the expected mass-flow error in response to external vibrations is obtained. The sensitivity to external vibrations is reduced, using passive and active vibration isolation solutions. Passive vibration isolation consists of a mass-spring-damper system between the floor and the measurement device. The performance is insufficient, because the suspension frequency is limited by the maximum stress in the connection tubes and a maximum allowable sag due to gravity. Therefore the passively suspended stage is extended with voice-coil actuators and absolute motion sensors to apply active vibration isolation control. Increased narrow-band vibration attenuation is achieved using loop shaping feedback control. Alternatively, an innovative feed-forward strategy is used, where by the parameters, which are tuned using an adaptive FxLMS technique, are based on the damping and stiffness properties of the stage. Based on the theoretic models, a proof of principle mechanism is realised for validation, whereby attenuation of more than 40 dB is achieved. The performance is limited by sensor noise and for the feedback solution also by the destabilising effect of higher order dynamics. The thesis presents a comprehensive analysis on how vibration isolation can be achieved. Narrow-band active vibration isolation control is successfully implemented in a design of a CMFM, for the first time.",
keywords = "METIS-310724, IR-96105",
author = "{van de Ridder}, Bert",
year = "2015",
month = "6",
day = "24",
doi = "10.3990/1.9789036538749",
language = "English",
isbn = "978-90-365-3874-9",
publisher = "University of Twente",
address = "Netherlands",
school = "University of Twente",

}

van de Ridder, B 2015, 'Vibration isolation for Coriolis Mass-Flow meters', University of Twente, Enschede. https://doi.org/10.3990/1.9789036538749

Vibration isolation for Coriolis Mass-Flow meters. / van de Ridder, Bert.

Enschede : University of Twente, 2015. 166 p.

Research output: ThesisPhD Thesis - Research UT, graduation UT

TY - THES

T1 - Vibration isolation for Coriolis Mass-Flow meters

AU - van de Ridder, Bert

PY - 2015/6/24

Y1 - 2015/6/24

N2 - A Coriolis Mass-Flow Meter (CMFM) is an active device based on the Coriolis force principle for direct mass-flow measurements, with high accuracy, range-ability and repeatability. The working principle of a CMFM is as follows: a fluid conveying tube is actuated to oscillate at a low amplitude. A fluid-flow in the vibrating tube induces Coriolis forces, which are proportional to the mass-flow, and affect the tube motion. Measuring the tube displacement in such a way that the change of its mode shape is determined, allows calculating the mass-flow. For low flows (< 1 kg/h), the Coriolis force induced motion is relatively small compared to motions induced by external vibrations, thus CMFMs designed to be sensitive to low flows are rather sensitive to external vibrations. Mainly external vibrations around the meter’s drive and Coriolis frequencies result in a measurement error. A model-based, quantitative estimation of the expected mass-flow error in response to external vibrations is obtained. The sensitivity to external vibrations is reduced, using passive and active vibration isolation solutions. Passive vibration isolation consists of a mass-spring-damper system between the floor and the measurement device. The performance is insufficient, because the suspension frequency is limited by the maximum stress in the connection tubes and a maximum allowable sag due to gravity. Therefore the passively suspended stage is extended with voice-coil actuators and absolute motion sensors to apply active vibration isolation control. Increased narrow-band vibration attenuation is achieved using loop shaping feedback control. Alternatively, an innovative feed-forward strategy is used, where by the parameters, which are tuned using an adaptive FxLMS technique, are based on the damping and stiffness properties of the stage. Based on the theoretic models, a proof of principle mechanism is realised for validation, whereby attenuation of more than 40 dB is achieved. The performance is limited by sensor noise and for the feedback solution also by the destabilising effect of higher order dynamics. The thesis presents a comprehensive analysis on how vibration isolation can be achieved. Narrow-band active vibration isolation control is successfully implemented in a design of a CMFM, for the first time.

AB - A Coriolis Mass-Flow Meter (CMFM) is an active device based on the Coriolis force principle for direct mass-flow measurements, with high accuracy, range-ability and repeatability. The working principle of a CMFM is as follows: a fluid conveying tube is actuated to oscillate at a low amplitude. A fluid-flow in the vibrating tube induces Coriolis forces, which are proportional to the mass-flow, and affect the tube motion. Measuring the tube displacement in such a way that the change of its mode shape is determined, allows calculating the mass-flow. For low flows (< 1 kg/h), the Coriolis force induced motion is relatively small compared to motions induced by external vibrations, thus CMFMs designed to be sensitive to low flows are rather sensitive to external vibrations. Mainly external vibrations around the meter’s drive and Coriolis frequencies result in a measurement error. A model-based, quantitative estimation of the expected mass-flow error in response to external vibrations is obtained. The sensitivity to external vibrations is reduced, using passive and active vibration isolation solutions. Passive vibration isolation consists of a mass-spring-damper system between the floor and the measurement device. The performance is insufficient, because the suspension frequency is limited by the maximum stress in the connection tubes and a maximum allowable sag due to gravity. Therefore the passively suspended stage is extended with voice-coil actuators and absolute motion sensors to apply active vibration isolation control. Increased narrow-band vibration attenuation is achieved using loop shaping feedback control. Alternatively, an innovative feed-forward strategy is used, where by the parameters, which are tuned using an adaptive FxLMS technique, are based on the damping and stiffness properties of the stage. Based on the theoretic models, a proof of principle mechanism is realised for validation, whereby attenuation of more than 40 dB is achieved. The performance is limited by sensor noise and for the feedback solution also by the destabilising effect of higher order dynamics. The thesis presents a comprehensive analysis on how vibration isolation can be achieved. Narrow-band active vibration isolation control is successfully implemented in a design of a CMFM, for the first time.

KW - METIS-310724

KW - IR-96105

U2 - 10.3990/1.9789036538749

DO - 10.3990/1.9789036538749

M3 - PhD Thesis - Research UT, graduation UT

SN - 978-90-365-3874-9

PB - University of Twente

CY - Enschede

ER -

van de Ridder B. Vibration isolation for Coriolis Mass-Flow meters. Enschede: University of Twente, 2015. 166 p. https://doi.org/10.3990/1.9789036538749