TY - JOUR
T1 - Wear-testing of a temporomandibular joint prosthesis
T2 - UHMWPE and PTFE against a metal ball, in water and in serum
AU - van Loon, J.-P.
AU - Verkerke, G.J.
AU - de Bont, L.G.M.
AU - Liem, R.S.B.
PY - 1999/8/1
Y1 - 1999/8/1
N2 - For a temporomandibular joint prosthesis, an estimation of the wear rate was needed, prior to patient application. Therefore, we determined the in vitro wear rate of the ball-socket articulation of this prosthesis, consisting of a metal head and an ultra-high molecular weight polyethylene (UHMWPE) cup. The basic testing configuration consisted of one 8-mm diameter stainless-steel ball. rotating between two conforming cups with a minimum thickness of 5 mm. For validation of the testing apparatus, two cup materials, in two lubricants, were tested. Both cup materials, UHMWPE and polytetrafluoroethylene (PTFE) were tested in deionized water, as well as in a serum-based solution. For UHMWPE in serum, eight samples were tested, for the other combinations four samples. For UHMWPE, the tests ran for 7 million cycles, for PTFE between 0.8 and 1.7 million cycles. For UHMWPE, the wear rate was 0.006 and 0.47 (mm3/106 cycles), in water and in serum, respectively. For PTFE, the wear rate was 2.8 and 47 (mm3/106 cycles), in water and in serum, respectively. For reason that testing in serum simulates the in vivo situation best, it was concluded that the wear rate of the TMJ prothesis articulation is 0.47 (mm3/106 cycles), which is considered acceptable.
AB - For a temporomandibular joint prosthesis, an estimation of the wear rate was needed, prior to patient application. Therefore, we determined the in vitro wear rate of the ball-socket articulation of this prosthesis, consisting of a metal head and an ultra-high molecular weight polyethylene (UHMWPE) cup. The basic testing configuration consisted of one 8-mm diameter stainless-steel ball. rotating between two conforming cups with a minimum thickness of 5 mm. For validation of the testing apparatus, two cup materials, in two lubricants, were tested. Both cup materials, UHMWPE and polytetrafluoroethylene (PTFE) were tested in deionized water, as well as in a serum-based solution. For UHMWPE in serum, eight samples were tested, for the other combinations four samples. For UHMWPE, the tests ran for 7 million cycles, for PTFE between 0.8 and 1.7 million cycles. For UHMWPE, the wear rate was 0.006 and 0.47 (mm3/106 cycles), in water and in serum, respectively. For PTFE, the wear rate was 2.8 and 47 (mm3/106 cycles), in water and in serum, respectively. For reason that testing in serum simulates the in vivo situation best, it was concluded that the wear rate of the TMJ prothesis articulation is 0.47 (mm3/106 cycles), which is considered acceptable.
KW - PTFE
KW - Temporomandibular joint
KW - Temporomandibular joint prosthesis
KW - UHMWPE
KW - Wear
KW - Wear testing
UR - http://www.scopus.com/inward/record.url?scp=0033178857&partnerID=8YFLogxK
U2 - 10.1016/S0142-9612(99)00042-3
DO - 10.1016/S0142-9612(99)00042-3
M3 - Article
C2 - 10458560
AN - SCOPUS:0033178857
SN - 0142-9612
VL - 20
SP - 1471
EP - 1478
JO - Biomaterials
JF - Biomaterials
IS - 16
ER -